Swissmedic rules for using the symbol
Put the name and address of your CH-REP directly next to the symbol.
The address must allow someone to contact the CH-REP. A P.O. box, an e-mail address or a phone number on its own is not enough.
There is no minimum size for the symbol or the name, but both must be clearly legible to the naked eye.
Instead of the symbol, you may write “CH authorised representative”, “CH-REP” or “Authorised representative for Switzerland”.
Where the CH-REP must appear
The required position depends on the type of device:
MDR devices: always on the label, and in the IFU (if applicable)
IVDR devices for self-testing: always on the label, and in the IFU (if applicable)
IVDR devices not for self-testing: always on the label, and in the IFU (if applicable), or in a document accompanying the device
Legacy devices (MDD, AIMDD, IVDD): depends on the device and whether an EU manufacturer or EC REP exists – see the Swissmedic information sheet
“Label” covers the device itself, the packaging of each unit and multi-unit packaging. An accompanying document can be a delivery note, invoice, customs document, sticker or the instructions for use.
The following label examples show correct and incorrect ways to state the CH-REP. They are simplified illustrations, not print templates.